15″ Multiparameter Patient Monitor MD9015T

Options
Invasive Blood Pressure (IBP)
Carbon dioxide (CO2) Module
Masimo SpO2 Module
Thermal Printer
Product Specification
Electrocardiogram (ECG)
Input mode
Five-lead ECG input / three-lead ECG input
Lead selection
I, II, III, aVR, aVL, aVF, V-
Gain selection
AUTO,0.25X,0.5X,1X,2X,4X
Operating mode
Diagnostic mode, monitoring mode, operation mode
Calibration signal
1mV ±5%
Protection
The isolation voltage 4000V, 50/60Hz. Anti-high-frequency interference of electric knife. Anti-defibrillation protection.
Pacemaker at the same time use
Support
Electrode off instructions
Each electrode (RL excluded)
Scanning speed
12.5mm/s, 25 mm/s, 50 mm/s
Common mode rejection ratio
≥90dB
Input impedance
≥5.0MΩ
Heart Raterate (HR)
Measuring range
Adult 30 ~ 300bpm, newborns 30 to 350 bpm
Resolution
1 bpm
Accuracy
< ±5bpm
Heart rate sensitivity
≥0.20mVp-p
Alarm range
15 to 350 bpm, upper and lower limits
Alarm indication
Audible and visual alarm, display the data and parameters flashing display alarm
message text prompts, there are three levels of alarm
Pulse Oximetry
SPO2 Range: 1% to 99%
Resolution: 1%
Accuracy: 70% to 99% range: ±2%;
1% to 69% range: ±5%;
Method: Dual wavelength LED
ST-segment
ST-segment measurement range
-2.0~+2.0mV
ST-segment measurement accuracy
Within the range of -0.8 ~ +0.8 mV: ± 0.02mV or reading± 10%,
Whichever is greater; Other range: undefined
Resolution
0.01mV
ST-segment update cycle
10s
ST segment alarm range
-2.0 ~ +2.0 MV, the upper and lower limits continuously adjustable

Arrhythmia analysis
Arrhythmia types
13 kinds of (scalable)
Respiration Rate
Rate: 2 to 120 bpm (CO2)
Resolution: 1 breath/min
NIBP (Non-invasive Blood Pressure)
Technique: Oscillometric during inflation
Range: Adult/Pediatric: 40 to 240 mmHg
Neonate: 10 to 150 mmHg
Measurement Cycle: < 40 sec. typical
Automatic Measurement
Cycles (Selectable): 2, 3, 5, 10, 15, 30 min; 1, 2, 4 hr
STAT Mode: 5 minutes of continuous readings
Max. Allowable Cuff
Pressure: Adult: 300mmHg
Pediatric: 240 mmHg;
Neonate: 150 mmHg
Resolution: 1 mmHg
Transducer Accuracy: ±2 mmHg or ±2% of reading,
Heart (Pulse) Rate
Range: 30 to 250 bpm
Accuracy: ±2 bpm or ±3%, whichever is greater